Weinmann OXYWAY Fix I Service Manual page 210

Pressure reducer
Table of Contents

Advertisement

Repairs to OXYWAY Click
307
307
210
EN
OXYWAY
Back
1. Lightly tension the cover clamping device 401 in the vise.
2. Insert the pressure reducer with O
3. Tighten the cover clamping device 401 in the vise.
4. Use the open-end wrench to loosen the inhalation outlet 301
5. Remove the inhalation outlet 301 from the multistep flow
301
6. Remove the adhesive residue from the threads on the multistep
17.7.2 Fitting the inhalation outlet
1. Pick up the inhalation outlet 301.
Risk due to adhesive in the gas flow!
If the first two threads of the inhalation outlet are moistened with
adhesive, this adhesive can get into the gas flow to the patient and
lead to airway irritation.
 Do not moisten the first two threads with adhesive.
2. Apply Loctite 413 to the thread of the inhalation outlet 301.
3. Insert the inhalation outlet 301 into the multistep flow selector 307.
301
4. Use the torque wrench to tightly screw in the inhalation
5. Loosen the vise.
Forward
cover clamping device 401.
When doing so, note:
• The O
sampling point 322 must be facing downwards.
2
• The cover clamping device 401 must clamp onto the housing
cover 308.
from the multistep flow selector 307.
selector 307.
flow selector 307.
When doing so, note:
• No dirt must get into the housing during cleaning.
• A brass brush must be used for the external thread.
• A thread tap must be used for the internal thread.
outlet 301 in the multistep flow selector 307.
When doing so, note: Loctite 413 must harden according to
the manufacturer's specifications.