14.28.1 Removing the O-ring from the housing cover
Requirement
• The multistep flow selector assembly 179 has been removed
• The spring for the piston 174 has been removed
• The piston 173 has been removed
• The multistep flow selector assembly 179 has been disassembled
1. Use the watchmaker's screwdriver 409 to carefully lift the
162
152
14.28.2 Fitting the O-ring in the housing cover
1. Clean the housing cover 162 with a dry cloth.
162
2. Use the Allen key to insert the O-ring 152 into the groove
152
3. Use the brush to apply a thin coat of oxygen-compatible
4. Assemble the multistep flow selector assembly 179
5. Fit the piston 173
6. Fit the spring for the piston 174
7. Fit the multistep flow selector assembly 179
Forward
(see "14.13.1 Removing the multistep flow selector assembly",
page
98).
(see "14.14.1 Removing the spring for the piston", page
(see "14.16.1 Removing the piston", page
(see "14.17.1 Disassembling the multistep flow selector assembly",
page
101).
O-ring 152 out of the housing cover 162.
in the housing cover 162.
lubricant 414 to the O-ring 152.
(see "14.17.2 Assembling the multistep flow selector assembly",
page
102).
(see "14.16.2 Fitting the piston", page
(see "14.14.2 Fitting the spring for the piston", page
(see "14.13.2 Fitting the multistep flow selector assembly",
page
98).