Pressure reducer
Table of Contents

Advertisement

Introduction
OXYWAY
Click
O2 sampling point
12
EN
OXYWAY
Back
The piston is acted on by a second spring which is adjustable by means
of a spindle system and which therefore generates different rated
outlet pressures. This means that the flow can be variably adjusted in
connection with the nozzle plate in the inhalation outlet. The adjusted
flow is read off at the flow indicator.
The pressure reducer is protected against excessive pressure by a safety
valve.
1.3.4 OXYWAY Click setup
Housing
Disposable
filter
Low pressure
Mechanics
During operation, the oxygen enters the housing from the central gas
supply system via the O
a pressure of 4.5 bar.
With OXYWAY Click, there is an orifice plate with various orifice
diameters in the multistep flow selector. The orifice diameter is
adjusted using the multistep flow selector. Different inhalation values
are output via the inhalation outlet depending on the setting of the
multistep flow selector.
Forward
Housing cover
sampling point on the pressure reducer with
2
Contents
Previous page
Multistep flow selector
Multistep
flow
selector
Orifice plate