• The multistep flow selector assembly 233 has been disassembled
(see "15.20.1 Disassembling the multistep flow selector assembly",
page
148).
1. Use the Allen key to remove the threaded pin 212 from the
multistep flow selector 211.
212
15.24.2 Fitting the threaded pin in the multistep flow
selector
1. Insert the threaded pin 212 into the multistep flow selector 211 and
212
screw in using the Allen key.
When doing so, note: The threaded pin 212 must not be screwed
into the hole for the orifice plate assembly 235.
2. Assemble the multistep flow selector assembly 233
(see "15.20.2 Assembling the multistep flow selector assembly",
page
149).
3. Fit the piston 226
4. Fit the spring for the piston 227
(see "15.17.2 Fitting the spring for the piston", page
5. Fit the multistep flow selector assembly 233
(see "15.16.2 Fitting the multistep flow selector assembly",
page
145).
15.25 Replacing the multistep flow selector
15.25.1 Removing the multistep flow selector
Requirement
• The multistep flow selector assembly 233 has been removed
(see "15.16.1 Removing the multistep flow selector assembly",
page
144).
• The spring for the piston 227 has been removed
(see "15.17.1 Removing the spring for the piston", page
• The piston 226 has been removed
(see "15.19.1 Removing the piston", page
• The multistep flow selector assembly 233 has been disassembled
(see "15.20.1 Disassembling the multistep flow selector assembly",
page
148).
Contents
Repairs to OXYWAY Fast II
(see "15.19.2 Fitting the piston", page
147).
OXYWAY
Previous page
148).
146).
146).