Weinmann OXYWAY Fix I Service Manual page 52

Pressure reducer
Table of Contents

Advertisement

Final inspection OXYWAY Fine I
500
510
52
EN
OXYWAY
Back
10.8 Checking the constant output (flow)
at the inhalation outlet
1. Screw the hand-screwed connection 273 onto the valve of
273
267
2. Screw the hand-screwed connection 273 tight by hand.
3. Close the multistep flow selector 267.
4. Screw the adapter 507 onto the inhalation outlet 257.
5. Screw the special kit 500 onto the adapter 507.
6. Connect the special kit 500 to the testing device 510.
7. Open the valve of the oxygen cylinder slowly.
8. Read off the cylinder pressure p1 at the contents indicator 250
9. Open the multistep flow selector 267 until the flow indicator 251
10.Read off the flow value at the testing device 510 and enter in
11.Open the multistep flow selector 267 until the flow indicator 251
12.Read off the flow value at the testing device 510 and enter in
13.Close the multistep flow selector 267 until the flow indicator 251
Forward
the oxygen cylinder.
and enter the value in the test report.
displays 8 l/min.
the test report.
Requirement:
The flow value is 8 l/min
±
displays 15 l/min.
the test report.
Requirement:
The flow value is 15 l/min
displays 3 l/min.
Contents
Previous page
0.8 l/min.
1.5 l/min.
±
Next page