Weinmann OXYWAY Fix I Service Manual page 55

Pressure reducer
Table of Contents

Advertisement

Back
Forward
11.2 Required test and auxiliary
equipment
• Leak detection spray 502
• Brush
• Cloth
• LPA cap, UNF 9/16" 504
• LPH adapter, G 3/8" internal thread - DIN coupling 13260 505
• LPA adapter, NF S 90-116 (AFNOR) coupling - G 3/8" internal
thread 506
• LPA adapter, G 3/8" internal thread - CGA V-5 external thread
(UNF 9/16") 507
• Set, supply test 501
• Special kit, ventilation test kit and pressure reducer Flow 500
• Oxygen cylinder 509
• Testing devices:
Flow measuring tube 0 to 20 l/min 510
Flow measuring tube 0 to 50 ml/min 512
Or
Ventilator tester
11.3 Preparing the final inspection
1. Enter the serial number of the pressure reducer in the test report.
2. Check the pressure reducer for external damage.
Requirement:
The pressure reducer shows no signs of external damage.
3. Check the label 321 on the pressure reducer
(see "18.5 Labels for pressure reducers", page
Requirement:
The label 321 on the pressure reducer is present and legible.
4. Check the connection thread on the pressure reducer.
To do so:
Screw the cap 504 onto the inhalation outlet 301.
Requirement:
The connection threads show no sign of external damage and
turn smoothly.