Operating Instruction STERIclave B 2200 / 2200 P
|
4 Installation
4.4 Preparations
CAUTION
+

4.4.2 Main switch for switching the STERIclave B on.

▶ Press the main switch to switch the STERIclave B on.
Note
The main switch should always be left switched on in order to provide a power
supply to the internal battery. Power consumption in standby mode is extremely
low.

4.4.3 Filling the waterways

Setting the elevation
Faulty sterilisation due to air pressure conditions
Correct sterilisation can only be assured if the STERIclave B is adapted to prevai‐
ling air pressure conditions/elevations.
▶ Set the elevation correctly.
▶ The settings will need to be modified if the STERIclave B moves to a different
elevation.
The factory setting specifies an elevation of 100 m. For heights between 0 and 200
metres, the setting does not have to be changed, as differences in elevation of ±100
m will not impair the operation of the unit.
For reliable sterilisation, the selected value must not deviate from the actual value
by more than 200 m. An incorrect setting will extend the duration of the vacuum
cycle and can trigger incorrect and/or premature AL8 or AL5 error messages.
See also: 8 Troubleshooting, Page 65
▶ Hold down the 1 button, and press the Power button.
"SET ALT 100 MT" will appear on the display.
▶ Press 1 to increase the current elevation.
▶ Press 2 to lower the current elevation.
▶ Press the Set button.
The set elevation is saved.
The water level will be topped up automatically.
"READY" will appear on the display to indicate that the filling process is com‐
plete.
▶ Open the door and wipe the chamber dry with a cloth.
The unit is ready for operation.