2   Pre-Installation - KaVo ORTHOPANTOMOGRAPH OP 3D Installation Manual

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

2   Pre-installation
• The CEPH upgrade kit consists of two packages, the cephalometric unit package and the upgrade
package. The upgrade package consists of mechanical parts needed for the upgrade, CEPH sensor
covers, installation screws, installation instructions, installation template, device manuals and
quality control tools (optional).
• This upgrade kit is applicable only to CEPH upgrade -ready devices, from serial number Kx1801417
onwards.
• CEPH upgrade kit requires device firmware FW 2.3 or newer.
• The upgrade kit is supplied with all required screws and EMC clamps. Always use the new screws
and clamps during the installation.
• The installation of CEPH upgrade kit varies between different product configurations.
• Make sure that the installation area has sufficient space for the CEPH arm and all CEPH movements.
Use the supplied installation template as a guide.
• CEPH upgrade may require you to move the Column in a different location. To move the column, the
carriage must be removed from the column. This task requires two persons to do.
• Drive the device down to a suitable and comfortable installation height before starting the
installation.
• Read all other pre-installation requirements on the OP 3D Installation manual, supplied with the
upgrade kit.
• The CEPH upgrade kit contains the CEPH arm and new linear movement mechanism and rotating
movement motor for the main device. The parts that need to be installed during the CEPH upgrade
depend on the device configuration and generation. See the table below to identify which parts
need to be installed.
PAN ONLY DEVICE
SN Kx1902594 or earlier
• Linear movement
mechanism
• Rotating movement motor
• CEPH arm
ORTHOPANTOMOGRAPH
SN Kx1902595 or newer
• CEPH arm
OP 3D
2   Pre-installation
3D DEVICE
Any SN
• CEPH arm
5