Instructions for use KEY Laser III 1243
3 Description of Instrument
|

3.5 Technical data and requirements

3.5 Technical data and requirements
Setup plan No.
Electrical lead
Mains power input line
Input voltage
By installing the "External voltage kit" or
an isolating transformer, the input volta‐
ge is expandable to
Frequency
Power consumption
Customer fuse protection 220-240 V
Customer fuse protection 100-127 V
Customer fuse protection 100-127 V
(Japan)
Classification
Device of protection class 1
Device not protected against the penet‐
ration of water
Type B application part
Operating conditions
Environmental temperature
Relative humidity
Condensate on unit
Suitable for operation up to 2,000 m above sea level.
Mode INT: Continuous operation with intermittent loading
(valid only for load on the laser unit. For the hand-piece mode, see hand-piece in‐
structions)
For a 1 min. pulse/1 min. pause with a unit setting of 600 mJ/10 Hz.
at an environmental temperature of 35
°C
at an environmental temperature of 25
°C
111 052 94 003
230 V Schuko socket
3 m long
230 V~
100 V~/110 V~/120 V~/127 V~
50/60 Hz
40-2200 W and 100-2500 VA
B16 or C16- automatic
25 A
20 A
15 – 35 °C
max. 85%
not permissible
permissible length of operation: 5 min.
permissible length of operation: 30 min.
33/99