Selecting And Starting A Program; Overview Of Programs - KaVo STERIclave B 2200 Operating Instruction

Table of Contents

Advertisement

Operating Instruction STERIclave B 2200 / 2200 P
|
6 Operation

6.4 Selecting and starting a program

Display
Program
LED 1
Sterilise 1
LED 2
Sterilise 2
LED 3
Fast cycle
LED 4
SPECIAL
S1
Prions
SPECIAL
P2
Maximum load
SPECIAL
P3
Maximum load
SPECIAL
P4
Individual
6.4 Selecting and starting a program

6.4.1 Overview of programs

Total cycle test
(Including drying)
18 min
36 min
15 min
32 min
20 min
38 min
15 to 80 min
Parameter
Load
(test according to DIN EN
13060:2004)
134 ℃
all sterilisation
5 min
items up to 134℃
3 pre-vacuum pha‐
for example: In‐
ses
struments + turbi‐
Drying 5 min (3
nes
min vacuum, 1 min
(Helix test)
ventilation)
121℃ 20 min
Solid sterilisation
3 pre-vacuum pha‐
items up to 121℃
ses
unpacked + single
Drying 11 min (4
packed
min vacuum, 2 min
(Helix test)
ventilation)
134℃ 4 min
Solid sterilisation
2 pre-vacuum pha‐
items up to 134℃
ses
unpacked
drying 3 min
(2 min vacuum, 1
min ventilation)
134 ℃
Against Creutz‐
19 min
feld-Jacob and
3 pre-vacuum pha‐
BSE pathogens*
ses
Drying: 3 + 1 min
134 ℃
Difficult materials
5 min
such as textiles
4 pre-vacuum pha‐
(Helix test)
ses
Drying: 3 + 1 min
121 ℃
Difficult sterilisati‐
20 min
on items
4 pre-vacuum pha‐
Up to 121°C
ses
(for example: sur‐
Drying: 4 + 2 min
gical cassettes)
Programmable by
Depending on the
the user:
set values
Temperature:
121-135 ℃
Time: 3-60 min.
Pre-vacuum pha‐
ses: 2.3 or 4
Drying with pre-va‐
cuum and ventila‐
tion: 2+1, 4+2 or
6+3 min
45/78
Type
(DIN EN 13060:2004)
B
B
S
B
B
B

Advertisement