Advertisement

4. Detach the cable from the circuit board below the upper cover half.
5. Press on both sides to release the lower cover half.
6. Pull the lower cover half off.

8.3 Replacing parts

8.3.1 Disconnecting and connecting the cables

NOTICE! Read this instruction before you start to disconnect any cables on the
device.
1. Check the number on the cable.
2. Check the number on the connector.
3. Connect the cable to the connector with the same number.

8.3.2 Replacing the PAN/3D sensor

NOTICE! Sensor head covers and rotator back and front covers need to be
removed, see chapters
the sensor head
Required tools: Allen key 3mm, adjustable spanner
ORTHOPANTOMOGRAPH
Removing the rotator back and front cover
covers.
OP 3D
8 Part replacement instructions
and
Removing
113