Instructions for use HealOzone 2130 C
|
7 Maintenance
7.2 Maintenance

7.2.2 Sensor/counter mode

Sensor mode
Sensor mode is the factory default. Sensor mode continuously checks the moisture
level and reports when the moisture level is too high and the desiccant is used up.
Treatment can continue after the desiccant is changed.
See also: 7.2.3 Replace desiccant, Page 48
Counter mode
Adverse environmental conditions and climatic influences can in certain circum‐
stances lead to frequent warnings regarding the moisture level.
If this should occur in rare instances, the user can change the counter mode.
The counter mode checks the operating hours (600 hours are programmed) and the
number of treatments (1000 treatments are programmed). If one of the programmed
values is exceeded, the warning message "Desiccant used up" appears
With this setting, the user can be sure that the device is fully functional without
exchanging the desiccant too early.
After the maximum operating hours or treatments are reached, the warning mes‐
sage "Desiccant used up" appears.
▶ Change desiccant.
See also: 7.2.3 Replace desiccant, Page 48
▶ Press SET button.
The device changes to the display "Ozone concentration can only be ensured once
the desiccant is changed."
▶ Press SET button.
The query "New desiccant" appears to confirm the desiccant change.
▶ Change the displayed value "No" to "Yes" with the PLUS and MINUS buttons.
▶ Press the SET button to confirm the desiccant change.
Note
If the desiccant change was not confirmed with "yes" in counting mode, the warning
message "Desiccant used up" appears after each treatment.