7. Press the M Key. The display at the left will appear.
bia5.1
8. Change the wiring as shown below.
5f165
l.cal
9. Wait until the count on Display No. 2 is sufficiently stable and then press
10. Press the M Key. The display at the left will appear. This display will not
str.i1
no
l.cal
11. Turn OFF the power to leave Calibration Mode.
Open (not connected).
STV
DMM
Short circuit.
ZERO-CON
OUTPUT
INPUT
Disconnect the STV and enable the thermocouple in the cold junction
compensator. Make sure that the STV is disconnected at this time.
the D Key. This tentatively registers the calibration data at this point.
appear if all of the required data has not been tentatively registered.
Press the U Key. Display No. 2 will show yes. Two seconds after the Key
is released or when the M Key is pressed, the tentatively registered
calibration data will be stored in EEPROM. If you do not wish to save the
data in EEPROM, press the M Key instead of the U Key.
• For a Controller with more than one input, connect as explained in step 2
and repeat steps 5 to 10.
• If a linear current output is selected, continue with the procedure in 9.6
Output Calibration (P. 9-12).
9.3 Thermocouple Input Calibration
Compensating leads of selected thermocouple.
However, compensating leads for a K
thermocouple can be used for E, R, S, B, and W
thermocouples.
9-7