Compatible SCPI Commands
Supported SCPI Commands
either be set via the center frequency and span, or via the start and stop frequency.
[SENSe<1|2>:]FREQuency:CENTer 0 to fmax
Specifies the analyzer's center frequency or the measuring frequency for span = 0.
Remote Command
Example
Remote Command Notes
Preset
State Saved
Min
Max
Backwards Compatibility SCPI
[SENSe<1|2>:]FREQuency:CENTer:STEP 0 to fmax
Specifies the center frequency's step width.
Remote Command
Example
Remote Command Notes
Preset
State Saved
Min
Max
Backwards Compatibility SCPI
[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK SPAN | RBW | OFF
Couples the center frequency's step width to the span (span >0) or to the resolution bandwidth (span =
0), or cancels the coupling.
Remote Command
138
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer?
FREQ:CENT 50MHz
FREQ:CENT?
FSE, FSP, FSU
fmax /2 with fmax = maximum frequency
Saved in instrument state.
Depends on instrument minimum frequency
Depends on instrument maximum frequency
:SENSe1:FREQuency:CENTer
[:SENSe]:FREQuency:CENTer:STEP <freq>
[:SENSe]:FREQuency:CENTer:STEP?
FREQ:CENT:STEP 120MHz
FSE, FSP, FSU
1MHz
Saved in instrument state.
1 Hz
27.0 GHz
:SENSe1:FREQuency:CENTer:STEP
[:SENSe]:FREQuency:CENTer:STEP:LINK SPAN | RBW | OFF
[:SENSe]:FREQuency:CENTer:STEP:LINK?