Rack Mount; Figure 2-3. Removing The Front Panel - Thermo Scientific 48iQ Instruction Manual

Carbon monoxide analyzer

Advertisement

Installation and Setup
Mounting Options

Rack Mount

2-4
48iQ Instruction Manual
Mounting in a rack includes removing the front panel and installing ears
and handles.
Equipment required:
Phillips drive, #2
1. Start by gripping from the top corners of the front panel and pull
outwards. See
Figure
Figure 2–3. Removing the Front Panel
2. Unfasten the four 8-32 pan head screws.
3. Slide ears outwards.
4. Use the same four 8-32 pan head screws to secure it.
5. Install the handles with the four 8-32 flat head screws that came with
the handle kit on the backside. See
2–3.
Figure
2–4.