Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries. AQUAfast, AQUASensors, BOD AutoEZ, ionplus, KNIpHE, LogR, No Cal, ORION, perpHect, PerpHecT, pHISA, pHuture, Pure Water, Sage, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS, and TURBO2 are registered trademarks of Thermo Fisher Scientific, Inc.
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meters Chapter 1 Introduction ............1 Chapter 2 Meter Overview.
Page 4
Chapter 8 Data Storage and Retrieval ..........25 Data Storage Settings .
Chapter 1 Introduction Thank you for your purchase of the Orion Star A211 benchtop pH meter, Star A221 portable pH meter or Star A321 portable pH meter. These meters are capable of measuring pH, mV, relative mV (RmV), ORP, and temperature in °C or °F.
Page 7
This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
(use a coin or your finger) and release the bottom portion of the battery cover. Orion Star A Series Benchtop Orion Star A Series Portable battery cover Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
COND POWER RS232 Star A221 and Star A321 Meter Back Panel Orion Star A211 benchtop meters only. Orion Star A211 benchtop meters only. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Orion Star A series benchtop meters include an electrode arm that can be attached to either side of the meter. 2. Choose the side of the meter to attach the arm. should be on the same side as the meter display. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Orion Star A211 and Star A321 meters only. Orion Star A211 and Star A321 meters only. is pressed. is pressed. sample ID is pressed. channel Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
(relative mV) and ORP. Orion Star A211 benchtop meters only. Press to start or stop the stirrer probe. Orion Star A211 benchtop meters only. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Orion Star A211 benchtop meters only. Meter Maintenance meter using the proper cleaning procedure for the type of spill. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
(Star A211 and Star A321 meters only) (Star A211 and Star A321 meters only) (Star A211 and Star A321 meters only) Temperature Calibration Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Menu used to select the resolution for the displayed measurement values. Menu used to select the buffer group used for automatic buffer recognition during pH calibrations. Buffer Group Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Page 16
Temperature Calibration built in to an electrode) to be adjusted by the operator. Menu used to select the source of the temperature measurement. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Menu used to set the language for all displayed meter prompts. Menu used to turn on or off the beep sound made every time a meter key is pressed. Alarm Beep Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
View Log Menu Storage and Retrieval chapter for detailed information on using this setup menu. Diagnostics Menu detailed information on using this setup menu. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Page 19
This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
(edit) to access the numeric entry screen, enter the slope value and press f2 (accept). f1 (meas) to proceed to the measurement mode or press f2 (print) to a printer or computer. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Press f2 (accept) to accept the entered mV value. f1 (meas) to proceed to the measurement mode or press f2 (print) the data to a printer or computer. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
(accept) once the value is stable. Press f1 (meas) to proceed to the measurement mode or press f2 (print) to proceed to the measurement mode and Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Page 23
This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Offset line. Press f2 (accept) f1 (meas) to proceed to the measurement mode or press f2 (print) to proceed to the measurement mode Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Page 25
This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Star A211 and Star A321 meters only – ready. Star A211 meters only to turn off the stirrer probe. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Page 27
This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Chapter 7 Methods The Orion Star A211 and Star A321 meters can save up to ten methods per channel and the corresponding calibration data for each method, for the fast and easy recollection of different meter setup parameters. The default meter Saving the Current Meter Settings as a New Method 2.
(accept) measurement mode. Delete is selected – press f2 (yes) to delete the method or press f3 (no) to keep the method. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
(HH), minutes (MM) or seconds (SS); press f3 (edit) to access the numeric entry screen; use the numeric entry screen to change the values and press f1 (back) to return to the measurement mode. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Data Log and press f3 (select). to highlight Off or On and press f3 (select). f1 (back) and then press to return to the measurement mode. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
(select). Press f2 (print) to print the data point or press f1 (back) to return to the list of data points. to return to the measurement mode. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
(enter) to select the highlighted vii. Press f2 (done) viii. Press f2 (accept) and then press f2 (yes) Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Log Clear and press f2 (accept). f2 (yes) 111111 and pressing f2 (accept) f1 (back) and then press to return to the measurement mode. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
A complimentary copy of general data logging software is available for use with the Orion Star A200 and Star A300 series meters. This can be found on the CD included with the meter or at www.thermoscientific.com/water. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Make sure that the electrode is connected to the appropriate meter input and the electrode Measurement Out of Range parameters are entered correctly in the setup menu. Clean the electrode according to the electrode Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
When the meter displays Keypad Test Passed, press f2 (next). Self Test Passed. Press to return to the measurement mode. Meter Checkout (pH) instructions. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Diagnostics and press f3 (select). 3. Press to highlight User Reset and press f2 (accept). f2 (yes) to reset the meter. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
About Meter and press f2 (accept). f1 (back) to return to the diagnostics menu or press to return to the measurement mode. Assistance Warranty and Registration Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Cat. No. Description STARA2110 STARA2115 STARA2120 STARA2125 STARA2126 STARA2130 STARA2135 STARA2136 STARA2140 STARA2145 STARA2146 Storage Solution STARA2147 STARA2148 STARA2150 STARA2155 Standard STARA2160 STARA2165 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Page 41
Star A321 pH Portable Meter STARA3215 STARA3220 Star A322 Conductivity Portable Meter STARA3225 STARA3230 STARA3235 STARA3240 STARA3245 STARA3250 STARA3255 Star A Meter Armor STARA3260 STARA3265 STARA3290 STARA3295 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Page 42
Stainless Steel ATC Probe 927007MD 9179BNMD 9180BNMD 9512HPBNWP 9609BNWP 8611BNWP 013005MD 013010MD 013020MD 013025MD 013016MD 083005MD 083010MD 083025MD 086030MD 087010MD 087020MD 087030MD 096019 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Page 43
This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
(de la class A) prescrites dans le Reglement sur le brouillage radioelectrique edicte par le WEEE Compliance directives, the recyclers in your country, and information on Thermo Scientific Orion products that may assist the Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Conforms with the following directives and standards: EN61326-1:2006 Electromagnetic Compatibility (EMC Directive) EN61010-1:2001 Safety Standards UL61010-1:2004 Safety requirements for electrical equipment for measurement, CAN/CSA C22.2 No. 61010-1-04 Singapore Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Warranty 3 year meter replacement Universal Power Adapter Operating Conditions Operating Ambient Temperature Operating Relative Humidity Storage Temperature Storage Relative Humidity Pollution Overvoltage Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
Page 47
Relative Accuracy ORP Mode Temperature Measurement Range Resolution Relative Accuracy Offset Calibration 1 point Data Logging Inputs ATC Probe Output Power AC Adapter Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
The electrode condition icon indicates the performance of the pH electrode, based on the last saved calibration and electrode measurement stability. Icon pH Electrode Definition of Icon troubleshoot the electrode. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...