Thermo Scientific Orion Star A211 Reference Manual

Thermo Scientific Orion Star A211 Reference Manual

Benchtop and portable ph meters
Hide thumbs Also See for Orion Star A211:
Table of Contents

Advertisement

Thermo Scientific Orion Star A211 Benchtop and
Star A221 and Star A321 Portable pH Meters
Reference Guide

Advertisement

Table of Contents
loading

Summary of Contents for Thermo Scientific Orion Star A211

  • Page 1 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meters Reference Guide...
  • Page 2 Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries. AQUAfast, AQUASensors, BOD AutoEZ, ionplus, KNIpHE, LogR, No Cal, ORION, perpHect, PerpHecT, pHISA, pHuture, Pure Water, Sage, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS, and TURBO2 are registered trademarks of Thermo Fisher Scientific, Inc.
  • Page 3: Table Of Contents

    Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meters Chapter 1 Introduction ............1 Chapter 2 Meter Overview.
  • Page 4 Chapter 8 Data Storage and Retrieval ..........25 Data Storage Settings .
  • Page 5 This page intentionally left blank.
  • Page 6: Chapter 1 Introduction

    Chapter 1 Introduction Thank you for your purchase of the Orion Star A211 benchtop pH meter, Star A221 portable pH meter or Star A321 portable pH meter. These meters are capable of measuring pH, mV, relative mV (RmV), ORP, and temperature in °C or °F.
  • Page 7 This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 8: Chapter 2 Meter Overview

    (use a coin or your finger) and release the bottom portion of the battery cover. Orion Star A Series Benchtop Orion Star A Series Portable battery cover Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 9: Electrodes And Other Connections

    COND POWER RS232 Star A221 and Star A321 Meter Back Panel Orion Star A211 benchtop meters only. Orion Star A211 benchtop meters only. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 10: Connecting Electrode Arm To The Benchtop Meter

    Orion Star A series benchtop meters include an electrode arm that can be attached to either side of the meter. 2. Choose the side of the meter to attach the arm. should be on the same side as the meter display. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 11: Meter Display

    Orion Star A211 and Star A321 meters only. Orion Star A211 and Star A321 meters only. is pressed. is pressed. sample ID is pressed. channel Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 12: Meter Keypad

    (relative mV) and ORP. Orion Star A211 benchtop meters only. Press to start or stop the stirrer probe. Orion Star A211 benchtop meters only. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 13: Meter Maintenance

    Orion Star A211 benchtop meters only. Meter Maintenance meter using the proper cleaning procedure for the type of spill. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 14: Chapter 3 Meter Setup

    (Star A211 and Star A321 meters only) (Star A211 and Star A321 meters only) (Star A211 and Star A321 meters only) Temperature Calibration Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 15: Ph Channel Menu

    Menu used to select the resolution for the displayed measurement values. Menu used to select the buffer group used for automatic buffer recognition during pH calibrations. Buffer Group Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 16 Temperature Calibration built in to an electrode) to be adjusted by the operator. Menu used to select the source of the temperature measurement. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 17: Settings Menu

    Menu used to set the language for all displayed meter prompts. Menu used to turn on or off the beep sound made every time a meter key is pressed. Alarm Beep Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 18: View Log Menu

    View Log Menu Storage and Retrieval chapter for detailed information on using this setup menu. Diagnostics Menu detailed information on using this setup menu. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 19 This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 20: Chapter 4 Ph, Rmv And Orp Calibration

    (edit) to access the numeric entry screen, enter the slope value and press f2 (accept). f1 (meas) to proceed to the measurement mode or press f2 (print) to a printer or computer. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 21: Calibration Editing

    Press f2 (accept) to accept the entered mV value. f1 (meas) to proceed to the measurement mode or press f2 (print) the data to a printer or computer. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 22: Orp Calibration

    (accept) once the value is stable. Press f1 (meas) to proceed to the measurement mode or press f2 (print) to proceed to the measurement mode and Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 23 This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 24: Chapter 5 Temperature Calibration

    Offset line. Press f2 (accept) f1 (meas) to proceed to the measurement mode or press f2 (print) to proceed to the measurement mode Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 25 This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 26: Chapter 6 Measurement

    Star A211 and Star A321 meters only – ready. Star A211 meters only to turn off the stirrer probe. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 27 This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 28: Chapter 7 Methods

    Chapter 7 Methods The Orion Star A211 and Star A321 meters can save up to ten methods per channel and the corresponding calibration data for each method, for the fast and easy recollection of different meter setup parameters. The default meter Saving the Current Meter Settings as a New Method 2.
  • Page 29: Editing, Copying Or Deleting An Existing Method

    (accept) measurement mode. Delete is selected – press f2 (yes) to delete the method or press f3 (no) to keep the method. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 30: Chapter 8 Data Storage And Retrieval

    (HH), minutes (MM) or seconds (SS); press f3 (edit) to access the numeric entry screen; use the numeric entry screen to change the values and press f1 (back) to return to the measurement mode. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 31: Computer And Printer Export Settings

    Data Log and press f3 (select). to highlight Off or On and press f3 (select). f1 (back) and then press to return to the measurement mode. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 32: Viewing The Calibration Log And Data Log

    (select). Press f2 (print) to print the data point or press f1 (back) to return to the list of data points. to return to the measurement mode. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 33: Exporting The Data Log To A Computer Or Printer

    (enter) to select the highlighted vii. Press f2 (done) viii. Press f2 (accept) and then press f2 (yes) Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 34: Clearing The Data Log

    Log Clear and press f2 (accept). f2 (yes) 111111 and pressing f2 (accept) f1 (back) and then press to return to the measurement mode. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 35: Interfacing A Printer

    A complimentary copy of general data logging software is available for use with the Orion Star A200 and Star A300 series meters. This can be found on the CD included with the meter or at www.thermoscientific.com/water. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 36: Chapter 9 Customer Services

    Make sure that the electrode is connected to the appropriate meter input and the electrode Measurement Out of Range parameters are entered correctly in the setup menu. Clean the electrode according to the electrode Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 37: Meter Self Test And Checkout

    When the meter displays Keypad Test Passed, press f2 (next). Self Test Passed. Press to return to the measurement mode. Meter Checkout (pH) instructions. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 38: Electrode Stability Test

    Diagnostics and press f3 (select). 3. Press to highlight User Reset and press f2 (accept). f2 (yes) to reset the meter. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 39: About Meter

    About Meter and press f2 (accept). f1 (back) to return to the diagnostics menu or press to return to the measurement mode. Assistance Warranty and Registration Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 40: Ordering Information

    Cat. No. Description STARA2110 STARA2115 STARA2120 STARA2125 STARA2126 STARA2130 STARA2135 STARA2136 STARA2140 STARA2145 STARA2146 Storage Solution STARA2147 STARA2148 STARA2150 STARA2155 Standard STARA2160 STARA2165 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 41 Star A321 pH Portable Meter STARA3215 STARA3220 Star A322 Conductivity Portable Meter STARA3225 STARA3230 STARA3235 STARA3240 STARA3245 STARA3250 STARA3255 Star A Meter Armor STARA3260 STARA3265 STARA3290 STARA3295 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 42 Stainless Steel ATC Probe 927007MD 9179BNMD 9180BNMD 9512HPBNWP 9609BNWP 8611BNWP 013005MD 013010MD 013020MD 013025MD 013016MD 083005MD 083010MD 083025MD 086030MD 087010MD 087020MD 087030MD 096019 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 43 This page intentionally left blank. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 44: Chapter 10 Meter Compliance

    (de la class A) prescrites dans le Reglement sur le brouillage radioelectrique edicte par le WEEE Compliance directives, the recyclers in your country, and information on Thermo Scientific Orion products that may assist the Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 45: Declaration Of Conformity

    Conforms with the following directives and standards: EN61326-1:2006 Electromagnetic Compatibility (EMC Directive) EN61010-1:2001 Safety Standards UL61010-1:2004 Safety requirements for electrical equipment for measurement, CAN/CSA C22.2 No. 61010-1-04 Singapore Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 46: Meter Specifications

    Warranty 3 year meter replacement Universal Power Adapter Operating Conditions Operating Ambient Temperature Operating Relative Humidity Storage Temperature Storage Relative Humidity Pollution Overvoltage Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 47 Relative Accuracy ORP Mode Temperature Measurement Range Resolution Relative Accuracy Offset Calibration 1 point Data Logging Inputs ATC Probe Output Power AC Adapter Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 48: Appendix 1 Advanced Meter Features

    Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 49: Electrode Condition Icon

    The electrode condition icon indicates the performance of the pH electrode, based on the last saved calibration and electrode measurement stability. Icon pH Electrode Definition of Icon troubleshoot the electrode. Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 50: Multipoint Calibration Curve Feature

    Multipoint Calibration Curve Feature Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide...
  • Page 51 This page intentionally left blank.
  • Page 52 Water Analysis Instruments North America 9 0 0 1 : 2 0 0 8 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 info.water@thermo.com Netherlands Tel: (31) 033-2463887 info.water.uk@thermo.com India Tel: (91) 22-4157-8800 wai.asia@thermofisher.com Japan Tel: (81) 045-453-9175 wai.asia@thermofisher.com China Tel: (86) 21-68654588...