First use
First use

Operation

Operation
CAUTION
Damage from soiled and moist cooling air.
Contaminated and moist cooling air can cause malfunctions and lead to
premature bearing wear.
▶ Make sure that the supply of cooling air is dry, clean and unconta‐
minated according to ISO 7494-2.
4 4 .1 C C heck the amount of water
CAUTION
Overheating of the tooth due to lack of cooling water.
Thermal damage to the dental pulp.
▶ Adjust the water amount for the spray cooling to a minimum of 50
cm
/min
3
5 5 O O peration
5.1 A A ttaching the medical device
WARNING
Release of the medical device during treatment.
A medical device that is not properly locked in place can release from the
motor coupling and fall off.
▶ Carefully pull on it before each treatment to ensure that the medical
device is securely locked on the motor coupling.
CAUTION
Damage
If the footswitch is pressed while you are connecting and removing the
medical device, you can damage the medical device and motor coupling.
▶ Do not connect or remove the medical device while pressing the
footswitch.
23
24
25
26