4 Operation
4.3 Laser tube/handpieces
4.3 Laser tube/handpieces
4.3.1 Laser tubing
Note
If the laser tube is removed when the unit is switched on, the safety switch ensures
that the KEY Laser 1243 cannot be switched to the Ready mode. Never allow the
laser tube to fall! Ensure that the laser tube couplings are not knocked against
anything and are never laid down roughly. Ensure that the exit window of the exit
coupling is not soiled or contaminated with fingerprints. This may damage the coa‐
ting and thus destroy the exit window
See also: Section 2.3 of handpiece instructions
▶ Push the laser tube coupling ② into the input coupling block ensuring correct
alignment of the plug.
▶ Screw down with lock nut.
▶ Guide the laser tube through the swing arm ①.
4.3.2 Coupling handpieces
Note
Ensure that the KEY Laser 1243 is switched to the standby mode.
38/99