Instructions for use ELECTROmatic M/C and PM/PC
8 Processing steps in accordance with DIN EN ISO 17664 | 8.2 Disinfection

8.1.4 Mechanical cleaning of the exterior and interior

Not applicable.

8.2 Disinfection

Growth of germs.
Infections.
▶ Before treating a patient, let the spray air and spray water exit for at least
20 seconds.
▶ Before the first start-up and after downtimes (weekends, holidays, vaca-
tions, etc.), the air and water lines must be purged and/or rinsed.
▶ The germ reduction of the ELECTROmatic must be carried out via the treat-
ment unit.
KaVo recommends the following products for germ reduction of the spray water
of the treatment unit:
▪ KaVo Oxygenal 6 made by KaVo Dental GmbH www.kavo.com
▪ BluTab made by ConFirm Monitoring Systems Inc www.blutab.com
®
▪ ICX
Water Treatment Tablets made by A-dec
▶ Use in accordance with manufacturer specifications.
Note
The unit must be disinfected manually only.

8.2.1 Manual external disinfection

▶ Use a soft disposable cloth and an approved disinfectant for germ reduction
by wiping down all visible surfaces of the unit, foot control surfaces, and
connecting cables. Make sure that all surfaces are fully wetted.
▶ Let the disinfectant act for the prescribed time.
▶ Dry the surfaces.
Permissible disinfectants (application range in accordance with the available
manufacturer's instructions for use and national guidelines. Please note material
safety data sheets.) KaVo recommends the following products based on the
compatibility of the materials. The microbiological efficacy must be ensured by
the disinfectant manufacturer.
▪ FD 322 (Dürr)
▪ Microcide AF Liquid (Schülke & Mayr)
▪ CaviCide (Metrex)
▪ Incidin Liquid (Ecolab)
Note
Comply with the instructions for use of the disinfectant.

8.2.2 Manual disinfection of the interior

The germ reduction of the ELECTROmatic must be carried out via the treatment
unit.
CAUTION