Manual Disinfection Of The The Exterior; Manual Disinfection Of The Interior; Mechanical Disinfection Of The Exterior And Interior; Care Products And Systems - Servicing - KaVo PROPHYflex 3 Instructions For Use Manual

2018 for sirona connection
Hide thumbs Also See for PROPHYflex 3:
Table of Contents

Advertisement

Instructions for use PROPHYflex 3 - 2018 for Sirona connection - 1.006.9920, - 1.006.9927
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing

7.3.1 Manual disinfection of the the exterior

KaVo recommends the following products based on compatibility of the materials. The
microbiological efficacy must be ensured by the disinfectant manufacturer.
▪ Mikrozid AF made by Schülke & Mayr (liquid or cloths)
▪ FD 322 made by Dürr
▪ CaviCide made by Metrex
Consumables required:
Cloths for wiping the medical device.
▶ Spray the disinfectant on a cloth, then wipe down the medical device and allow
the disinfectant to act according to the instructions of the
disinfectant manufacturer.
▶ Follow the instructions for use of the disinfectant.

7.3.2 Manual disinfection of the interior

The efficacy of manual internal disinfection must be demonstrated by the manufac-
turer of the disinfection agent. With KaVo products, use only disinfection agents that
have been released by KaVo with respect to the compatibility of materials (e.g. WL-
cid / made by ALPRO).
▶ Follow the instructions for use of the disinfectant.

7.3.3 Mechanical disinfection of the exterior and interior

KaVo recommends washer disinfectors in compliance with EN ISO 15883-1 that are
operated with alkaline cleaning agents.
The validations were conducted with the VARIO-TD program, the cleaning agent
neodisher® MediClean and the neutralisation agent neodisher® Z.
▶ For program settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the thermodisinfector.
▶ In order to prevent damage to the medical device due to residual fluid, make sure
that the inside and outside of the device is dry after the end of the cycle. Remove
any residual liquids from the interior and exterior of the medical device using com-
pressed air.
The drying procedure is normally part of the cleaning program of the thermodisin-
fector.
Note
Please comply with the instructions for use of the washer disinfector.

7.4 Care products and systems - Servicing

Improper care.
Malfunction or property damage.
▶ Do not clean the PROPHYflex with oils or care spray.
NOTICE
21 / 26