Beam Alignment Test; Ground Test; Motor Movements; Position Detectors - KaVo ORTHOPANTOMOGAPH OP 2D Service Manual

Digital panoramic x-ray unit
Table of Contents

Advertisement

Beam alignment test

Ground test

Motor movements

Position detectors

ORTHOPANTOMOGAPH OP 2D
Check the beam alignment. Refer to the Installation
manual of the unit for information on how to do this.
Disconnect the unit from the main power supply before
carrying out this test.
For ME EQUIPMENT with an APPLIANCE INLET the
impedance between the earth pin in the APPLIANCE
INLET and any part that is PROTECTIVELY EARTHED
shall not exceed 0.1 ohm. The grounding resistance is
measured between APPLIANCE INLET ground pin and
any metal part of the unit.
The resistance MUST be <0.1 ohm.
Switch the unit off and then manually rotate the rotating
unit to check that the stepper motor moves freely and
without any looseness.
Switch the unit on, activate the T-mode "test" and then
take an exposure to check that the motors operate
smoothly and without any excessive noise.
Press the up/down keys to check the Z-motor (vertical
carriage movement). The motor must operate smoothly
and without any excessive noise.
Press the up key and drive the unit up. Make sure that
the unit stops moving at its uppermost position.
Press the down key and drive the unit down. Make sure
that the unit stops moving at its lowermost position.
Manually rotate the rotating unit to one of its end posi-
tions and then press the return button and make sure
that the rotating unit returns to the ready position. Re-
peat the test for the other end position.
Use service command optotest, see section 5.2 Ser-
vice Commands, to check the rotating unit optosensors.
Manually turn the rotating unit in one direction and then
press return. Repeat for the other direction.