Once the scale(s) have been calibrated in this manner, the
touch screen will return to the scale check screen and display
the hang weights prompt, indicating the scale needs to go through
the 500 and 1000 gram checks again. Scale checks will need to
be performed on those scales that did not initially pass.
Typical Scale Check Screen
Continue to perform the scale checks using the 500 and
1000 gram weights as outlined previously.
If, at any time during this process, the scale fails its check, begin
the scale calibration sequence again. If the scale continues to
fail calibration, call a qualified service representative for
assistance.
07-19-01-518 – January 2009
3-17