07-19-01-518 – January 2009
A
C
LYX
OMPONENT
C
HAPTER
23. Place the donor and anticoagulant lines securely in the
channels directly in front of the donor safety clamp. The air
detectors are located within the left channel.
Lines Placed in Air Detectors
WARNING: The air detector housing and tubing channel
must be dry for proper function.
24. Check all tubings to ensure they are clear of the cassette
door latches.
25. Close the cassette compartment door and ensure it is secure.
WARNING: Do not operate the instrument if the cassette
compartment door is open. Do not operate the instrument if
the cassette compartment or pump block gasket is damaged.
26. Hang the in-process container on the in-process scale. The
icon on the container matches the icon just above the scale.
27. Hang the red cell container on the red cell scale. The icon
on the container matches the icon just above the scale.
28. Hang the plasma container on the plasma scale. The icon on
the container matches the icon just above the scale.
Note: Ensure the containers hang freely on the scale hangers so
they do not touch the instrument or the other containers. Ensure
the tubings are not twisted.
C
S
OLLECTION
YSTEM
4C – A
2RBC-NLR C
LYX
O
'
M
PERATOR
S
ANUAL
P
OLLECTION
ROCEDURE
4C-19