Download Print this page

Fenwal 4R5707 Operator's Manual page 268

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
4B – A
RBC/P
HAPTER
LYX
Check Button
4B-44
S
O
YSTEM
PERATOR
C
P
LASMA
OLLECTION
Jumbo Plasma Product
1.
Seal the tubing between the cassette and the jumbo plasma
product container.*
Note: *This tubing may be used to connect extra storage
container(s) with the use of a sterile connect device, if desired.
Ensure an adequate length of tubing is left attached to the red cell
product container and/or the jumbo product container for this
purpose.
2.
If the plasma is to be kept as a jumbo unit, follow the next
two steps to remove air from the unit, if desired.
3.
To remove air from the jumbo unit, close the clamp on one
satellite container.
4.
Invert the jumbo unit, open its clamp, and express the air
into the second satellite container. Close the clamp on the
second satellite container when finished.
5.
Seal the line on the jumbo unit next to the port. Separate the
satellite containers from the kit, remove the clamp and
discard.
Note: Do not seal the tubing (step 5), if a split plasma product
configuration was selected, and the plasma has not been split into
the satellite containers. Follow the directions for sealing under
Splitting the Plasma Product.
Note: The lines to the ACD, saline, and preservation solution
containers must be sealed at this time if these containers will be
disposed as non-biohazard waste.
6.
Verify that all seals have been made, then touch the
check button.
WARNING: Handle and dispose of all biologically
contaminated materials according to local and national
government requirements.
WARNING: Disposable kits and sterile solutions are for
single-use only.
'
M
S
ANUAL
ROCEDURE
07-19-01-518 – January 2009

Advertisement

loading

This manual is also suitable for:

4r5725