Target Volume Button
Check Button
07-19-01-518 – January 2009
A
C
LYX
OMPONENT
Typical Collection Volume Overlay
33. Touch the "360" target volume button on the overlay.
Note: If instrument defaults have been changed or if using
another procedure, choose the lowest collection volume
available.
34. Touch the check button on the overlay.
Typical Donor Data Screen Complete
Note: The instrument will not allow entry of donor or product
inputs inconsistent with configured settings of the instrument's
Administrative Settings. An audible tone and flashing data
button indicates entry of unacceptable data input.
35. Verify the clamp to the donor line and sampling pack are
closed.
C
S
OLLECTION
YSTEM
C
3 – O
HAPTER
PERATING CONSIDERATIONS
(Test Mode)
(Test Mode)
O
'
M
PERATOR
S
ANUAL
3-37