Fenwal 4R5707 Operator's Manual page 328

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
4D– 2RBC S
HAPTER
Check Button
Unload Kit Prompt
4D-10
S
O
YSTEM
PERATOR
U
R
INGLE
NIT
ECOVERY
Note: The lines to the ACD, saline, and preservation solution
containers must be sealed at this time if these containers will be
disposed as non-biohazard waste.
3.
Touch the check button to verify the kit has been sealed.
Removing the Kit
The remove kit screen will be displayed on the touch screen.
Typical Procedure Information Screen with Remove Kit
1.
Remove the red cell product from the scale and label the
product according to applicable standard operating
procedures.
Note: Fenwal recommends performing sterile transfer of the
contents of the Red Cell Container to one of the Red Cell Storage
Containers from the kit, or another approved red cell storage
container, by use of a sterile connecting device. Any excess air
can be expelled into the Red Cell Container. The process and
product information should be documented according to facility
standard operating procedures.
Refer to Chapter 4A A
"Removing the Kit" for instructions regarding kit removal.
Properly dispose of the kit according to standard operating
procedures.
2.
Touch the check button to verify the kit has been removed
from the instrument.
'
M
S
ANUAL
P
ROCEDURE
Prompt (SUR)
2RBC-LR Collection Procedure,
LYX
07-19-01-518 – January 2009

Advertisement

loading