Fenwal 4R5707 Operator's Manual page 371

Alyx component collection system

Advertisement

Code
Icons
1 -
Non-Recoverable
Alarm Screen
1002
3007
None
3008 -
Scale has been
Disturbed
3012
4001
Check
Venipuncture
Open Clamp
Check for Kink in
Line
4002
Check
Venipuncture
Check for Kink in
Line
4003 -
Check for Kink in
Line
4008
4010
Non-Recoverable
Alarm Screen
4011 -
Check for Kink in
Line
4013
07-19-01-518 – January 2009
A
LYX COMPONENT COLLECTION
Meaning
The system has detected a software
error.
The system has detected an issue
with the weigh scales.
The system has detected an issue
with the weigh scale highlighted on
the screen.
The system has detected that the
fluid flow from the donor is
insufficient.
The clam p on the donor line may be
closed.
The tubing lines may be kinked.
The system has detected that the
fluid flow to the donor is
insufficient.
The needle may need adjustment
during a return phase.
The tubing lines may be kinked.
The system has detected a fluid flow
issue in the part of the kit
highlighted on the screen.
The system has detected a hardware
calibration failure.
The system has detected a fluid flow
issue in the part of the kit
highlighted on the screen.
S
YSTEM
C
HAPTER
Resolution
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
- If problem persists, contact the Hotline.
- Check the scale that is highlighted on the
screen.
- Resume the procedure.
- If problem persists, contact the Hotline.
- Check donor line for kinks or occlusions.
- Check needle placement.
- Check donor's arm.
- Encourage donor to squeeze.
- Decrease draw/return rate.
- Check the donor line; ensure that the clamp is
open.
- Check for kinks in the tubing to and from the
filter.
- Check for kinks in the tubing from the RBC
container to the cassette.
- Check for kinks in the tubing from the
preservation solution container to the cassette.
- Resume the procedure.
- Check needle placement.
- Check donor's arm.
- Check donor line for kinks or occlusions.
- Decrease return rate.
- Check for kinks in the tubing to and from the
filter.
- Check for kinks in the tubing from the red cell
container to the cassette.
- Check for kinks in the tubing from the
preservation fluid container to the cassette.
- Check donor line; ensure the clamp is open.
- Check donor line for kinks or occlusions.
- Check clamps on storage container lines to
ensure they are open.
- Resume the procedure.
- Check for kinks in the tubing that is highlighted
on the screen.
- Resume the procedure.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
- Check for kinks in the tubing that is highlighted
on the screen.
- Resume the procedure.
O
'
M
PERATOR
S
ANUAL
5 – T
ROUBLESHOOTING
5-21