Place the tubing elbow into the optical sensor, ensuring it is
flush in the slot, and press the end of the umbilicus into the
umbilicus holder.
Tubing Elbow and Umbilicus Properly
20.
Lift the kit tray to close the centrifuge door by pushing down
on the door until the handle latches.
WARNING: Do not operate the instrument if the centrifuge
door is open or damaged.
21.
Remove the remaining containers and the donor line from
the kit tray and set them aside.
22.
Open the cassette compartment door, remove the cassette
from the kit tray, and discard the tray.
WARNING: If the kit tubing or cassette is damaged,
discontinue the procedure.
23.
Align the tabs on the top of the cassette with the cassette
guides on the instrument.
ANUAL
Installed
07-19-01-518 – January 2009