Fenwal 4R5707 Operator's Manual page 249

Alyx component collection system

Advertisement

Break Cannulae
Prompt
07-19-01-518 – January 2009
A
C
LYX
OMPONENT
C
HAPTER
When the kit checks are complete, the touch screen will
display the break cannulae prompt and the Kit Timer
flashing 60:00 (minutes and seconds).
Typical Screen with Break Cannulae Prompt
Note: The operator has the option to delay starting prime.
Refer to Chapter 3, Automatic Cassette Refresh and Kit Timer
for more information.
4.
Break the cannulae on all solution containers (saline (if
available), red cell preservation solution, and anticoagulant –
in any order). The cannulae are located at the inlet of each
solution container. To completely break a cannula, grasp it
in both hands, snap it at a 90-degree angle in one direction,
and then bend it in the opposite direction. To ensure proper
fluid flow, the cannula should be snapped as close to the
middle of the breakaway portion as possible.
Breaking a Cannula
C
S
OLLECTION
4B – A
RBC/P
LYX
LASMA
O
'
M
YSTEM
PERATOR
S
C
P
OLLECTION
ROCEDURE
ANUAL