Note: The instrument should always be transported with the
leveling feet adjusted to ensure the instrument is in the lowest
position possible.
When conducting mobile applications, follow the complete
instrument instructions for use.
The following information may be used for the installation of the
A
Instrument on a mobile coach.
LYX
Instrument Dimensions (See Physical Specifications)
Mounting Specifications
The diagram below shows the locations and tolerances for the
four feet on the A
instrument.
LYX
For permanent mounting: From the center of each foot, a 1.25
inches (31.7 mm) diameter circle pattern is drawn to contain two
clearance holes (0.375 inches [9.5 mm] diameter). The holes
must be oriented 180° apart from each other, but their
orientation to the other feet are not important. Two - 2 1/8
inches (54 mm) long (1/4-20; must be US threads) Phillips head
screws are required for attachment to each foot. Screws of the
appropriate length must be used such that they do not interfere
with the leveling mechanism on the bottom of each foot.