Fenwal 4R5707 Operator's Manual page 182

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
4A – A
2RBC-LR C
HAPTER
LYX
Loading the
Kit
Load Kit Prompt
4A-12
S
O
'
M
YSTEM
PERATOR
S
P
OLLECTION
ROCEDURE
The following directions are recommended for the installation of
the disposable kit on the instrument.
WARNING: Use only Fenwal approved kits on the
A
Component Collection System.
LYX
WARNING: Disposable kits and sterile solutions are for
single use only.
WARNING: Discard the kit and needle if the needle
becomes contaminated.
1.
Ensure the is displayed on the touch screen.
2.
Open the centrifuge door.
3.
Remove the shrink-wrap from the kit.
Note: The shrink-wrap does not function as a sterile barrier. The
needle cover serves as the sterile barrier and should remain intact
and undisturbed until use.
4.
Remove the clip holding the solutions in place, if applicable.
5.
Raise the kit support on the right side of the instrument and
place the kit tray on the kit support.
Kit Placed on Instrument Using Kit Support
WARNING: Use only clear solutions.
WARNING: If fluid or particulate matter is present inside
the tubing lines, cassettes, or storage containers, do not use
the disposable kit.
ANUAL
07-19-01-518 – January 2009