07-19-01-518 – January 2009
A
C
LYX
OMPONENT
C
HAPTER
17 Place the tubing elbow into the optical sensor, ensuring it is
flush in the slot, and press the end of the umbilicus into the
umbilicus holder.
Tubing Elbow and Umbilicus Properly Installed
18. Lift the kit tray to close the centrifuge door by pushing down
on the door until the handle latches.
WARNING: Do not operate the instrument if the
centrifuge door is open or damaged.
19. Remove the remaining containers and the donor line from
the kit tray and set them aside.
20. Open the cassette compartment door, remove the cassette
from the kit tray, and discard the tray.
WARNING: If the kit tubing or cassette is damaged,
discontinue the procedure.
21. Align the tabs on the top of the cassette with the cassette
guides on the instrument.
C
S
OLLECTION
YSTEM
4C – A
2RBC-NLR C
LYX
O
'
M
PERATOR
S
ANUAL
P
OLLECTION
ROCEDURE
Optical
Sensor
Umbilicus
Holder
4C-17