When all fluid has been returned to the donor, an audible tone
will sound and the disconnect donor prompt will appear during the
final red cell collection. The donor may be disconnected at this
time, even though the final red cell collection may not be
complete. The remaining red cells in the separation chamber will
be pumped into the red cell container to collect the final targeted
red cell volume.
Typical Main Collection Screen - Disconnect Donor
To disconnect the donor, follow these steps:
1.
Close the clamp on the donor line.
2.
Remove the pressure cuff from the donor's arm.
3.
Remove the needle while simultaneously sliding it into the
needle protector. Ensure the needle is completely covered
by the needle protector.
4.
Apply pressure to the venipuncture site to stop blood flow.
5.
Follow institutional guidelines for venipuncture site care and
release the donor according to applicable standard operating
procedures.
'
M
S
ANUAL
ROCEDURE
07-19-01-518 – January 2009