Fenwal 4R5707 Operator's Manual page 187

Alyx component collection system

Advertisement

07-19-01-518 – January 2009
A
C
LYX
OMPONENT
C
HAPTER
12.
With the umbilicus pointing toward the front of the
instrument, slide the separation chamber into the carrier.
Ensure that the rim of the separation chamber fits snugly in
the carrier mount.
13.
Close the latch to lock the separation chamber into position.
Push down on the latch to ensure it is completely closed.
Pull up on the latch once to verify that the latch has locked
into place. If closed correctly, the latch cannot be pulled
back up.
Note: If the latch does pull up, reseat the separation chamber into
the carrier. Then close as instructed in step 13.
14.
Bend the umbilicus upward and insert it within the umbilicus
guides on the carrier. Ensure the umbilicus is secure within
the guides.
Umbilicus
Guides
Separation Chamber Properly Installed in Carrier
C
S
OLLECTION
YSTEM
4A – A
2RBC-LR C
LYX
O
'
M
PERATOR
S
ANUAL
P
OLLECTION
ROCEDURE
Latch
4A-17