CAUTION: Do not place liquids on top of a closed instrument.
CAUTION: Ensure that no devices other than the pressure cuff
and barcode scanner (if used) are attached to the connection
ports on the back of the A
connected to the A
LYX
CAUTION: The instrument should be plugged in and powered-
up at least once every six months for 24 hours to maintain battery
charge should the instrument be stored for extended periods of
time. Not doing so will cause the battery to become functionally
inoperative and no longer able to hold an electrical charge, which
will necessitate battery replacement.
CAUTION: Ensure the caps from the back of the A
instrument connection pressure ports 1 and 2 are removed before
use. These caps must be removed from the ports prior to
connecting the pressure cuff and operating the instrument.
Leaving the caps in place may result in alarms and possible
service calls.
ANUAL
instrument while a donor is