Fenwal 4R5707 Operator's Manual page 264

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
4B – A
RBC/P
HAPTER
LYX
Check Button
4B-40
S
O
YSTEM
PERATOR
C
P
LASMA
OLLECTION
WARNING: If an alarm or interruption occurs while the
system is processing the red cell product after procedure
completion, a small amount of blood components may enter
the red cell preservation solution container. Dispose of all
potentially biohazardous materials per standard operating
procedures.
CAUTION: Red blood cells collected from some donors may
have extended filtration times resulting in the potential for
ineffective filtration and leukoreduction. The final product (after
filtration) should be evaluated for residual leukocytes and red cell
quality any time a reduction of flow through the leukoreduction
filter occurs during the automated leukoreduction filtration
process. The Procedure Information Screen will display a
message if the product needs to be checked.
CAUTION: If the operator terminates the leukoreduction
process, an adequate amount of red cell preservation solution
may not have been added to the red cell product to support
42-day storage. Disposition of the product should be left to the
medical director.
Typical Procedure Information Screen
When the processing phase is complete, the check button will be
displayed. The operator should touch the check button after the
donor has been appropriately disconnected.
'
M
S
ANUAL
ROCEDURE
with Filtration Time
07-19-01-518 – January 2009