Fenwal 4R5707 Operator's Manual page 81

Alyx component collection system

Advertisement

07-19-01-518 – January 2009
A
C
LYX
OMPONENT
Follow the steps below to complete a scale calibration:
1.
Place a 1000 gram weight on the scale indicated by the
highlighted button for calibrating at 1000 grams.
Typical Scale Calibration with 1000 grams
2.
Touch the highlighted 1000 gram calibration button to verify
the 1000 gram weight has been properly placed on the
scale. The button animation will indicate the instrument is
calibrating the scale to 1000 grams.
Typical Scale Calibrating to 1000 grams
C
S
O
OLLECTION
YSTEM
C
3 – O
HAPTER
PERATING CONSIDERATIONS
'
M
PERATOR
S
ANUAL
3-15