Ensure the kit has been properly loaded.
WARNING: If fluid or particulate matter is present inside
the tubing lines, cassette or storage containers do not use the
disposable kit.
WARNING: Do not remove the needle from its protective
cover until it is time to perform venipuncture.
CAUTION: If the disposable kit is not primed, it should be used
within 8 hours after being loaded on the instrument. However,
once the cannulae have been broken on a closed kit, or once kit-
checks have started on a functionally closed kit, a procedure must
begin within 1 hour. After this time, the instrument will alarm and
prompt the operator to remove the kit. The operator must press
the STOP button, end the procedure and remove the kit. A new
procedure may then be started. (See Chapter 3, Automatic Cassette
Refresh and Kit Timer for additional information.)
C
OLLECTION
4A – A
2RBC-LR C
LYX
Kit Properly Loaded
S
O
'
M
YSTEM
PERATOR
S
P
OLLECTION
ROCEDURE
ANUAL
4A-23