Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
2 – S
O
HAPTER
YSTEM
2-22
S
O
YSTEM
PERATOR
VERVIEW
During the collection procedure, the instrument continuously
monitors the weight of the containers on these scales along with
the saline and anticoagulant scales. The software in the
instrument automatically converts the weight on each scale into
fluid volumes based on specific gravity. This allows the
instrument to monitor fluid flow rates, extracorporeal volume,
solution usage, and leaks within the system.
CAUTION: The system monitors the weight of containers
hanging on the scales. Do not lift or pull on the containers as
this may cause alarms.
'