07-19-01-518 – January 2009
A
C
LYX
OMPONENT
C
HAPTER
Note: Use the appropriate directions (found below) for
installation of the filter depending on the type of filter fixture
available.
8.
Place the leukoreduction filter in the slot inside the single
leukoreduction filter fixture and snap the fixture closed.
Single Slot Fixture
9.
Place the leukoreduction filter in the larger, inside slot of the
dual leukoreduction filter fixture (the correct slot will have
two red cell icons at the top of the filter), and snap the
fixture closed.
Ensure the holder is secure on the instrument and the tubing is not
pinched or constricted.
Note: The leukoreduction filter fixture should remain closed once
the filter has been installed. The fixture should not be opened until
it is time to burp the final red cell product. If the leukoreduction
filter fixture is discovered to be in the open position any time prior
to product processing, simply close the fixture and continue with
the procedure.
Note: Failure to place the leukoreduction filter into the correct
slot of the dual filter fixture may compromise the product.
10.
Leave the paper strip on the red cell storage container
segment tubing, if desired, and hang the container on the
storage container hanger located on the front of the
instrument. Ensure that the clamps to the red cell storage
containers are open.
C
S
OLLECTION
YSTEM
4A – A
2RBC-LR C
LYX
Dual Slot Fixture
O
'
M
PERATOR
S
ANUAL
P
OLLECTION
ROCEDURE
4A-15