Download Print this page

Fenwal 4R5707 Operator's Manual page 217

Alyx component collection system

Advertisement

Seal Kit Prompt
Check Button
07-19-01-518 – January 2009
A
C
LYX
OMPONENT
C
HAPTER
Sealing the Kit
After the operator verifies that the red cell product is burped, the
Procedure Information Screen will display the seal kit prompt.
Typical Procedure Information Screen with
To seal the kit, perform the following steps:
1.
Seal the line between the leukoreduction filter and "Y"
junction, immediately next to the "Y."
2.
Separate the red cell storage containers from the rest of the
kit.
3.
Seal lines attached to the solution containers, if desired.
Note: The lines to the ACD, saline, and preservation solution
containers must be sealed at this time if these containers will be
disposed as non-biohazard waste.
4.
Touch the check button to verify the kit has been sealed and
the red cell storage containers have been removed.
WARNING: Handle and dispose of all biologically
contaminated materials according to local and national
government requirements.
WARNING: Disposable kits and sterile solutions are for
single-use only.
C
S
OLLECTION
YSTEM
4A – A
2RBC-LR C
LYX
Seal Kit Prompt
O
'
M
PERATOR
S
ANUAL
P
OLLECTION
ROCEDURE
4A-47

Advertisement

loading

This manual is also suitable for:

4r5725