Fenwal 4R5707 Operator's Manual page 373

Alyx component collection system

Advertisement

Code
Icons
7011 -
Non-Recoverable
Alarm Screen
Remove Kit
7014
7015
Check for Kink in
Line
Check Cannulae
Check Container
Loading
7016
Check for Kink in
Line
Scale has been
Disturbed
7018
Non-Recoverable
Alarm Screen
7020
Scale has been
Disturbed
7025
Air in Line
Air in AC Line
07-19-01-518 – January 2009
A
LYX COMPONENT COLLECTION
Meaning
The system has detected an issue
during AC prime.
The system has detected an issue
during AC prime.
The system has detected a fluid flow
issue in the part of the kit
highlighted on the screen.
The system has detected an issue
with the disposable kit.
The system has detected an
unexpected weight on the scales.
The system has detected air in the
donor line.
S
YSTEM
C
HAPTER
Resolution
- Press the STOP button to end the procedure.
- Remove the kit.
- Check to ensure the air detector is dry.
- Start a new procedure with a different kit.
- Check that the AC container is properly
installed on the AC scale.
- Check that the AC cannula is broken.
- Check the AC line for kinks or occlusions.
- Resume the procedure.
- Check for kinks in the tubing that is highlighted
on the screen.
- Resume the procedure.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
- Check the container on the scale that is
highlighted on the screen.
- Check the tubing leading to that container for
potential twisting.
-Check the appropriate kit has been loaded for
the procedure.
- Resume the procedure.
- Purge the air from the line.
- Check the donor line for air.
- Ensure the AC container is placed correctly on
the AC scale.
- Ensure the line from the anticoagulant
container to the cassette has no air.
- Ensure the anticoagulant line from the cassette
to the "Y" junction on the donor line has no air.
- If air is found, purge the air again. If no air is
found and the check button becomes active,
resume the procedure.
Note: If the air continues to persist after
several attempts to clear air from the line,
discontinue the procedure. Do not return
fluids.
Note: Small air bubbles in the AC line
dislodged from cassette surfaces will be drawn
into the kit during normal operation of the
instrument.
Note: Too many attempts to clear the air may
result in an IVD alarm (10504).
O
'
M
PERATOR
S
ANUAL
5 – T