Fenwal 4R5707 Operator's Manual page 420

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
7 – S
S
HAPTER
YSTEM
Considerations for
Mobile Use
7-14
S
O
YSTEM
PERATOR
PECIFICATIONS
Battery Specifications
Rechargeable nickel metal hydride battery cells power the
instrument's back-up battery system.
Battery Support Time
The back-up battery can support standby mode for
approximately 30 minutes.
Battery Recharge
The back-up battery system is recharged to full power within
10 hours or less when the instrument is connected to a primary
power source. Operating the instrument does not interrupt the
recharging process.
CAUTION: The instrument should be plugged in and powered-
up at least once every six months for 24 hours to maintain battery
charge should the instrument be stored for extended periods of
time. Not doing so will cause the battery to become functionally
inoperative and no longer able to hold an electrical charge, which
will necessitate battery replacement.
Note: The instrument must be plugged into a primary power
source and turned ON for the battery to recharge.
Battery Replacement or Disposal
For information on replacing or disposing of the back-up battery,
contact a qualified service representative.
Note: The back-up battery does not contain cadmium, mercury,
lead, or latex.
The A
LYX
mobile use. The instrument may be used for blood collections at
fixed or mobile sites. Care should be used when transporting the
instrument to avoid damage.
Note: The instrument should be turned OFF and closed before
it is relocated. Scale checks will be necessary after relocation of
the instrument.
The A
LYX
using a Fenwal approved transport cart and case or in the original
shipping carton. Do not drop, knock, or drag the instrument as
this may cause damage.
'
M
S
ANUAL
Component Collection System has been designed for
instrument should be transported from site to site
07-19-01-518 – January 2009