42. Touch the start button on the screen to verify the clamp on
the donor line has been opened. Collection will begin.
Typical Main Collection Screen –
Donor Draw Phase (Test Mode)
43. Follow institutional guidelines to determine whether the
sampling pack will remain with or be separated from the kit.
Operational checks do not require samples to be taken.
44. During a draw phase, verify that the donor flow indicator is
displayed with green segments. Completely occlude the
donor line. Verify the alarm tones sound and an alarm
screen is displayed.
45. Touch the mute button on the screen and verify the alarm
tones are no longer heard.
46. Touch the check button to acknowledge the alarm.
47. Remove occlusion from donor line.
48. Touch the start button to resume the procedure.
49. While in a draw phase, lift the anticoagulant container off the
anticoagulant scale for at least 10 seconds. Verify the alarm
tone sounds and an alarm screen is displayed.
CAUTION: Do not operate the instrument if the centrifuge
area exhibits extreme or unusual noise.
C
S
O
OLLECTION
YSTEM
C
3 – O
HAPTER
PERATING CONSIDERATIONS
'
M
PERATOR
S
ANUAL
3-39