Fenwal 4R5707 Operator's Manual page 234

Alyx component collection system

Advertisement

A
C
LYX
OMPONENT
C
4B – A
HAPTER
LYX
Loading the Kit
Load Kit Prompt
4B-10
C
S
O
OLLECTION
YSTEM
RBC/P
C
LASMA
OLLECTION
The following directions are recommended for the installation of
the disposable kit on the instrument.
WARNING: Use only Fenwal approved kits on the A
Component Collection System.
WARNING: Disposable kits and sterile solutions are for
single use only.
WARNING: Discard the kit and needle if the needle
becomes contaminated.
1.
Ensure the load kit prompt is displayed on the touch screen.
2.
Open the centrifuge door.
3.
Remove the shrink-wrap from the kit.
Note: The shrink-wrap does not function as a sterile barrier.
The needle cover serves as the sterile barrier and should remain
intact and undisturbed until use.
4.
Remove the clip holding the solutions in place, if applicable.
5.
Raise the kit support on the right side of the instrument and
place the kit tray on the kit support.
WARNING: Use only clear solutions.
'
M
PERATOR
S
ANUAL
P
ROCEDURE
Kit Placed on Instrument Using Kit Support
LYX
07-19-01-518 – January 2009