Fenwal 4R5707 Operator's Manual page 236

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
4B – A
RBC/P
HAPTER
LYX
Red Cell
Anticoagulant
Preservation
Icon
Solution Icon
4B-12
S
O
YSTEM
PERATOR
C
P
LASMA
OLLECTION
Saline Container on the Saline Scale
8.
Place the leukoreduction filter in the smaller, outside slot of
the dual leukoreduction filter fixture (the correct slot will
have one red cell icon at the top of the filter) and snap the
fixture closed. Ensure the holder is secure on the instrument
and the tubing is not pinched or constricted.
Leukoreduction Filter Properly Installed in Dual
Note: The dual leukoreduction filter fixture should remain
closed once the filter has been installed. The fixture should not
be opened until it is time to burp the final red cell product. If the
dual leukoreduction filter fixture is discovered to be in the open
position any time prior to product processing, simply close the
fixture and continue with the procedure.
'
M
S
ANUAL
ROCEDURE
Leukoreduction Filter Fixture
07-19-01-518 – January 2009
Slot for
1 unit
RBC
filter