Fenwal 4R5707 Operator's Manual page 330

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
C
4D– 2RBC S
HAPTER
2RBC-NLR Single
Unit Recovery
(SUR) Procedure –
Closed Kit
Ending the
Procedure
4D-12
S
O
OLLECTION
YSTEM
U
R
INGLE
NIT
ECOVERY
The operator ends the procedure by touching the STOP button,
and selects the option to Process product (if available). The
remaining red cells from the in-process container and the
separation chamber will be processed and pumped to the red cell
container, if needed. Red cell preservation solution will
automatically be added and the mixture will be pumped to the red
cell storage container.
Any collected red cells that exceed the maximum Single Unit
Recovery volume, but are less than the minimum absolute red cell
volume for a 2RBC collection, will be pumped into the in-process
container or returned to the donor, if the ability to return is an
option.
1. Touch the STOP button to end the procedure.
An overlay screen similar to the one displayed below will be
shown with two possible options:
• Return fluid to donor
• Do not return fluid to donor
2.
Close the clamp on one of the red cell storage containers.
3.
Touch the Do not return fluid to donor button if no fluid return is
desired.
The Process product and Do not process product overlay will be
displayed similar to the one shown below.
'
M
PERATOR
S
ANUAL
P
ROCEDURE
Typical End Procedure Screen Overlay
07-19-01-518 – January 2009