Alyx component collection system

Advertisement

Audio Check
Prompt
07-19-01-518 – January 2009
A
C
LYX
OMPONENT
Performing Daily Set-Up
The following items will be required in order to perform the
initial daily checks on the A
One 500 gram calibrated weight with hanger
One 1000 gram calibrated weight - or - Two 500 gram
calibrated weights with hanger(s)
1.
Push the STOP button to activate a check of the
functionality of the button. If the instrument does not
respond when the STOP button is pushed, contact your
qualified service representative. Verify the audio check
prompt is displayed, the signal light on the top of the
instrument becomes solid green, and an audible tone
sounds. If the tone cannot be heard, contact your
qualified service representative. Refer to the following
screen.
Typical A
LYX
C
S
OLLECTION
YSTEM
C
3 – O
HAPTER
PERATING CONSIDERATIONS
instrument:
LYX
Logo Screen with Audio
Check Prompt
O
'
M
PERATOR
S
ANUAL
3-11