Nomogram for
A
2RBC
LYX
Procedures
07-19-01-518 – January 2009
A
C
LYX
OMPONENT
Appendix
The table below indicates the maximum blood donation volumes
cleared by the FDA cleared by the FDA for collection of two
units of red blood cells using the A
procedure.
Donor
Donor
Weight
Height
Male
≥ 5' 1"
130–149 lbs.
≥ 5' 1"
150–174 lbs.
≥ 5' 1"
175 lbs. and
over
Female
≥ 5' 5"
150–174 lbs.
≥ 5' 5"
175 lbs. and
over
*Pre-donation hematocrit may be determined using a 5 minute spun hematocrit or
equivalent quantitative method.
The tables below indicate the maximum red cell donation
volumes used in Europe for the collection of two units of red
cells using the 2RBC-LR procedure. The following nomograms
have been developed for the A
edition of the "Guidelines to the preparation, use, and quality
assurance of blood components", Council of Europe Publishing
and ensured to conform to the 14
be configured to use the nomogram accepted by regulatory
agencies in your country. To change the nomogram in use, refer
to the Administrator's Guide for the A
Collection System or contact your service representative.
C
S
OLLECTION
YSTEM
2RBC-LR or 2RBC-NLR
LYX
Donor
Donor
Hematocrit*
Hemoglobin
≥ 40 %
≥ 13.3 g/dL
≥ 40 %
≥ 13.3 g/dL
≥ 40 %
≥ 13.3 g/dL
≥ 40 %
≥ 13.3 g/dL
≥ 40 %
≥ 13.3 g/dL
instrument using the 11
LYX
th
edition. The instrument will
Component
LYX
O
'
M
PERATOR
S
ANUAL
A
PPENDIX
Maximum
Total
Absolute
RBC
Volume
360 mL
400 mL
420 mL
360 mL
400 mL
th
A-3

Advertisement