Fenwal 4R5707 Operator's Manual page 192

Alyx component collection system

Advertisement

A
C
C
LYX
OMPONENT
OLLECTION
C
4A – A
2RBC-LR C
HAPTER
LYX
4A-22
S
O
'
M
YSTEM
PERATOR
S
P
OLLECTION
ROCEDURE
27.
Close the cassette compartment door and ensure it is
secured.
WARNING: Do not operate the instrument if the cassette
compartment door is open. Do not operate the instrument if
the cassette compartment or pump block gasket is damaged.
28.
Hang the in-process container on the in-process scale. The
icon on the container matches the icon just above the scale.
29.
Hang the red cell container on the red cell scale. The icon on
the container matches the icon just above the scale.
30.
Hang the plasma container on the plasma scale. The icon on
the container matches the icon just above the scale.
Note: Ensure the containers hang freely on the scale hangers so
they do not touch the instrument or the other containers. Ensure
the tubings are not twisted.
Containers Properly Installed on
31.
Remove the paper strip from the donor line and set aside.
Note: Do not place the needle and donor line on the
anticoagulant scale or on top of the anticoagulant container.
Placing anything other than the anticoagulant container on the
anticoagulant scale may cause alarms.
32.
Close the clamps on the donor line and the line leading to the
sampling pack.
ANUAL
Hanging Scales
07-19-01-518 – January 2009